HN-OSC10, 8P is used for pressure monitoring and control for automatic sequential analysis of up to 8 canister air samples. Each channel of OSC8 is equipped with a digital pressure sensor, a nitrogen valve and a nitrogen filling line connected to a common nitrogen inlet. The pressure value of each canister is shown on the screen mounted on the front panel. By switching on the nitrogen valve, the operator can pressurize the corresponding canister with nitrogen if the pressure of the canister is too low for analysis.
Main parameters:
Canister connection: VICI® 1/16 inch female connector X 8
Sampling connection:VICI® 1/16 inch female connector X 8
Nitrogen connection: Swagelok® 1/8 inch bulkheads tube fitting
Pressure range: 0 ~ 100 psia
Szie: 320 mm (L) x 380 mm (H) x 220 mm (d)